Uploaded image for project: 'dcm4chee'
  1. DCMEE-1056

C-FIND RSP always contains Other Patient ID Sequence if PIX Query is configured

    Details

    • Type: Bug
    • Status: Open
    • Priority: Minor
    • Resolution: Unresolved
    • Affects Version/s: dcm4chee-2.10.15, dcm4chee-2.11.0, dcm4chee-2.12.0, dcm4chee-2.13.0, dcm4chee-2.13.1, dcm4chee-2.13.2, dcm4chee-2.13.3, dcm4chee-2.13.4, dcm4chee-2.13.5, dcm4chee-2.13.6, dcm4chee-2.14.0, dcm4chee-2.14.1, dcm4chee-2.14.2
    • Fix Version/s: None
    • Component/s: DICOM
    • Labels:
      None

      Description

      C-FIND RSP contains Other Patient ID Sequence if PIX Query is configured, also if the original C-FIND RQ did not contain a return key for it.
      As a consequence, the client may include the Other Patient ID Sequence in created Evidence Documents (e.g. Secondary Capture images) stored to the archive, which causes the archive to append specified Other Patient IDs to the Patient Record associated with the Study, to which the Evidence Document is stored (s. DCMEE-158 Closed ). Such internal stored linkage of Patient IDs may become out-of-sync with the Patient ID linkage managed by the external PIX manager.

        Attachments

          Activity

            People

            • Assignee:
              gunter Gunter Zeilinger
              Reporter:
              gunter Gunter Zeilinger
            • Votes:
              0 Vote for this issue
              Watchers:
              0 Start watching this issue

              Dates

              • Created:
                Updated: